# Self-gated fat-suppressed cardiac cine MRI.

Ingle RR, Santos JM, Overall WR, McConnell MV, Hu BS, & Nishimura DG.

_Magn Reson Med. 73(5):1764-74. doi: 10.1002/mrm.25291. Epub 2014 May 7._

2015-05-01

## Abstract

PURPOSE: To develop a self-gated alternating repetition time balanced steady-state free precession (ATR-SSFP) pulse sequence for fat-suppressed cardiac cine imaging.

METHODS: Cardiac gating is computed retrospectively using acquired magnetic resonance self-gating data, enabling cine imaging without the need for electrocardiogram (ECG) gating. Modification of the slice-select rephasing gradients of an ATR-SSFP sequence enables the acquisition of a one-dimensional self-gating readout during the unused short repetition time (TR). Self-gating readouts are acquired during every TR of segmented, breath-held cardiac scans. A template-matching algorithm is designed to compute cardiac trigger points from the self-gating signals, and these trigger points are used for retrospective cine reconstruction. The proposed approach is compared with ECG-gated ATR-SSFP and balanced steady-state free precession in 10 volunteers and five patients.

RESULTS: The difference of ECG and self-gating trigger times has a variability of 13 ± 11 ms (mean ± SD). Qualitative reviewer scoring and ranking indicate no statistically significant differences (P > 0.05) between self-gated and ECG-gated ATR-SSFP images. Quantitative blood-myocardial border sharpness is not significantly different among self-gated ATR-SSFP ( 0.61±0.15 mm -1), ECG-gated ATR-SSFP ( 0.61±0.15 mm -1), or conventional ECG-gated balanced steady-state free precession cine MRI ( 0.59±0.15 mm -1).

CONCLUSION: The proposed self-gated ATR-SSFP sequence enables fat-suppressed cardiac cine imaging at 1.5 T without the need for ECG gating and without decreasing the imaging efficiency of ATR-SSFP.

[PubMed abstract](http://www.ncbi.nlm.nih.gov/pubmed/24806049)

![Bob S Hu, M.D.](https://dyn.mktgcdn.com/p/OLtyV-80yDbOmh0GcQI3xnNzxEtgLP1RmJcZvvKP3u4/fit=crop,gravity=auto,dpr=2,width=188,height=250,quality=100)

Bob S Hu, M.D.

Principal investigator

[View profile](https://www.sutterhealth.org/research/investigators/bob-s-hu-md-1049681033)

## Research topics

-   [Cardiovascular Diseases](https://www.sutterhealth.org/research-search-results?query=%22%5C%22Cardiovascular+Diseases%5C%22%22)

## Related publications

[View all publications](https://www.sutterhealth.org/research-search-results?verticalKey=research_publication)

-   Publication
    
    ### [No beneficial effects of pine bark extract on cardiovascular disease risk factors](https://www.sutterhealth.org/research/publications/pine-bark-extract-on-cardiovascular-disease-1055312556)
    
    Drieling RL, Gardner CD, Ma J, Ahn DK, Stafford RS.
    
    Arch Intern Med. 170(17):1541-7. doi: 10.1001/archinternmed.2010.310.
    
    2010 Sep 27
    
-   Publication
    
    ### [Home blood pressure monitoring among adults—American Heart Association Cardiovascular Health Consumer Survey, 2012](https://www.sutterhealth.org/research/publications/home-blood-pressure-monitoring-among-adults-1055312805)
    
    The Journal of Clinical Hypertension
    
    2017 Jun 19
    
-   Publication
    
    ### [Commentary on Costello JM, et al. Predictions by clinicians of quality of life for children and adolescents with cardiac disease.](https://www.sutterhealth.org/research/publications/commentary-predictions-qol-childhd-cardiac-disease-1055312358)
    
    Patel M, Wilson SR.
    
    In: Cabana M
    
    ed. Year Book of Pediatrics 2016. Philadelphia
    
    PA: Elsevier; 2015: 199-202. ISBN: 978-0-323-44293-0.
    
    2015 Dec 01
    
-   Publication
    
    ### [Association of the cardiometabolic staging system with individual engagement and quality of life in the US adult population.](https://www.sutterhealth.org/research/publications/cardiometabolic-staging-ind-engagemt-qol-us-adults-1055312307)
    
    Tsai SA, Xiao L, Lv N, Liu Y, Ma J.
    
    Obesity (Silver Spring). doi: 10.1002/oby.21907. \[Epub ahead of print\]
    
    2017 Jul 16
    
-   Publication
    
    ### [Combined outer volume suppression and T2 preparation sequence for coronary angiography.](https://www.sutterhealth.org/research/publications/outer-vol-suppressn-t2-prepseq-ht-angiography-1055312448)
    
    Luo J, Addy NO, Ingle RR, Hargreaves BA, Hu BS, Nishimura DG, Shin T.
    
    Magn Reson Med. 74(6):1632-9. doi: 10.1002/mrm.25575. Epub 2014 Dec 17.
    
    2015 Dec 01
    

[View all publications](https://www.sutterhealth.org/research-search-results?verticalKey=research_publication)

```json
{"@context":"https://schema.org","@type":"MedicalOrganization","name":"Sutter Health","@id":"https://www.sutterhealth.org/#medicalorganization","url":"https://www.sutterhealth.org","logo":"https://www.sutterhealth.org/static/images/sutter-health-logo.svg","address":{"@type":"PostalAddress","streetAddress":"2200 River Plaza Drive","addressLocality":"Sacramento","addressRegion":"CA","postalCode":"95833","addressCountry":"US"},"areaServed":{"@type":"AdministrativeArea","name":"California"},"sameAs":["https://www.facebook.com/SutterHealth","https://www.instagram.com/sutterhealth/","https://x.com/SutterHealth","https://www.youtube.com/sutterhealth","https://www.linkedin.com/company/sutter-health","https://en.wikipedia.org/wiki/Sutter_Health"]}
```

```json
{"@context":"https://schema.org","@type":"BreadcrumbList","@id":"https://www.sutterhealth.org/research/publications/self-gated-fat-suppressed-cardiac-cine-mri-1055312491#breadcrumblist","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https://www.sutterhealth.org/"},{"@type":"ListItem","position":2,"name":"Research","item":"https://www.sutterhealth.org/research"},{"@type":"ListItem","position":3,"name":"Research Publications","item":"https://www.sutterhealth.org/research/publications"},{"@type":"ListItem","position":4,"name":"Self-gated fat-suppressed cardiac cine MRI."}]}
```

```json
{"@context":"https://schema.org","@type":"MedicalScholarlyArticle","@id":"https://www.sutterhealth.org/research/publications/self-gated-fat-suppressed-cardiac-cine-mri-1055312491#medicalscholarlyarticle","name":"Self-gated fat-suppressed cardiac cine MRI.","headline":"Self-gated fat-suppressed cardiac cine MRI.","abstract":"PURPOSE: To develop a self-gated alternating repetition time balanced steady-state free precession (ATR-SSFP) pulse sequence for fat-suppressed cardiac cine imaging. METHODS: Cardiac gating is computed retrospectively using acquired magnetic resonance self-gating data, enabling cine imaging without the need for electrocardiogram (ECG) gating. Modification of the slice-select rephasing gradients of an ATR-SSFP sequence enables the acquisition of a one-dimensional self-gating readout during the unused short repetition time (TR). Self-gating readouts are acquired during every TR of segmented, breath-held cardiac scans. A template-matching algorithm is designed to compute cardiac trigger points from the self-gating signals, and these trigger points are used for retrospective cine reconstruction. The proposed approach is compared with ECG-gated ATR-SSFP and balanced steady-state free precession in 10 volunteers and five patients. RESULTS: The difference of ECG and self-gating trigger times has a variability of 13 ± 11 ms (mean ± SD). Qualitative reviewer scoring and ranking indicate no statistically significant differences (P > 0.05) between self-gated and ECG-gated ATR-SSFP images. Quantitative blood-myocardial border sharpness is not significantly different among self-gated ATR-SSFP ( 0.61±0.15 mm -1), ECG-gated ATR-SSFP ( 0.61±0.15 mm -1), or conventional ECG-gated balanced steady-state free precession cine MRI ( 0.59±0.15 mm -1). CONCLUSION: The proposed self-gated ATR-SSFP sequence enables fat-suppressed cardiac cine imaging at 1.5 T without the need for ECG gating and without decreasing the imaging efficiency of ATR-SSFP.","description":"The proposed self-gated ATR-SSFP sequence enables fat-suppressed cardiac cine imaging at 1.5 T without the need for ECG gating and without decreasing the imaging efficiency of ATR-SSFP.","keywords":["Cardiovascular Diseases"],"datePublished":"2015-05-01","url":"https://www.sutterhealth.org/research/publications/self-gated-fat-suppressed-cardiac-cine-mri-1055312491","inLanguage":"en-US","identifier":[{"@type":"PropertyValue","propertyID":"PMID","value":"24806049"},{"@type":"PropertyValue","propertyID":"DOI","value":"10.1002/mrm.25291"}],"sameAs":["https://pubmed.ncbi.nlm.nih.gov/24806049/","https://doi.org/10.1002/mrm.25291"],"isPartOf":{"@type":"Periodical","name":"Magn Reson Med"},"author":[{"@type":"Person","name":"Ingle RR"},{"@type":"Person","name":"Santos JM"},{"@type":"Person","name":"Overall WR"},{"@type":"Person","name":"McConnell MV"},{"@type":"Person","@id":"https://www.sutterhealth.org/research/investigators/bob-s-hu-md-1049681033","name":"Hu BS","url":"https://www.sutterhealth.org/research/investigators/bob-s-hu-md-1049681033"},{"@type":"Person","name":"Nishimura DG"}],"sourceOrganization":{"@type":"MedicalOrganization","name":"Palo Alto Medical Foundation Research Institute"},"about":[{"@type":"DefinedTerm","name":"Cardiovascular Diseases","url":"https://www.sutterhealth.org/research-search-results?facets=%5B%7B%22i%22%3A%22c_researchTopics.name%22%2C%22v%22%3A%5B%22Cardiovascular+Diseases%22%5D%7D%5D&verticalKey=research_publication"}],"mainEntityOfPage":{"@type":"WebPage","@id":"https://www.sutterhealth.org/research/publications/self-gated-fat-suppressed-cardiac-cine-mri-1055312491","breadcrumb":{"@id":"https://www.sutterhealth.org/research/publications/self-gated-fat-suppressed-cardiac-cine-mri-1055312491#breadcrumblist"}}}
```